D5 = 10 ms (power switching delay) S3 = 3H (power at which the 90 degree pulse will be measured) D1 = 1.5s (relaxation delay) P1 = 2.8us (this value will be varied to find the 90) RD = 0 PW = 0 NS = 1 DS = 0Use the AU command to execute the pulse program (it will automatically perform the EFP command) and go into EP mode and the water peak with the A (Additive phase correction) command. Check to see that your phasing is okay by typing AU again. The spectrum should be properly phased, if not, repeat the above steps. Next vary P1 by small increments (5 us) to find the 180 or 360 pulse (spectrum will be null). Divide the value by 2 or 4 to get the 90 pulse. You can also adjust your SW at this point if desired.
Now you are ready to begin setting up the TOCSY experiment. If you have a small molecule
with positive NOE's (MW<500-600) use the DANTOC.AU=D1
pulse program, or if you have a
larger molecule with negative NOE's, use the DANCTOC.AU=D1
pulse program.
NOTE AS OF 1/2/96: DANTOC.AU WAS TESTED AND THERE APPEARS TO BE AN ERROR IN THE PHASE CYCLING!! USE THIS PULSE PROGRAM AT YOUR OWN RISK! PERSONALLY I RECOMMEND USING DANCTOC.AU FOR ALL MOLECULES AT THIS TIME!!!AS
Use your measured P1 value to calculate your loop counter L1 to get an appropriate
mixing time for the TOCSY experiment. The mixing time should be 0.1/JHH, but for
molecules with 3 to 4 component spin systems mixing times of 50-100 msec are usually
appropriate. Lean towards 100 msec if your spin systems are long and/or your coupling
constants are small. If you are using the DANCTOC.AU pulse program, your loop counter should be:
L1 = (mixing time - 1msec)/147.2*p1Don't forget to express p1 in milliseconds!
D20 = 2.6*p1Or, if you are using the DANTOC.AU pulse program, calculate your loop counter as follows:
L1 = (mixing time - 1msec)/64.67*p1Again, p1 should be in milliseconds.
Now you are ready to begin setting up the TOCSY experiment. If you have a small molecule
with positive NOE's (MW<500-600) use the DANTOC.AU=D1 pulse program, or
if you have a larger molecule with negative NOE's, use the DANCTOC.AU=D1 pulse program.
Now AS the appropriate experiment and set the parameters as follows:
D2 = 30 ms S1 = 15H P8 = 25 us D8 = 250 us L0 = 6000The above parameters control the DANTE water suppression. If the suppression is too high or low, adjust L0 to lenghten or decrease the saturation time, or increase or decrease the power, S1. Be careful of increasing the power too much! The parameters for the rest of the TOCSY experiment are as follows:
D7 = 30 ms S2 = 3H P4 = 180 degree pulse measured P1 = 90 degree pulse measured D0 = 3 us P0 = 1 ms P3 = 90 degree pulse measured D20 = set as calculated above (only for DANCTOC.AU experiment) P5 = 90 degree pulse measured P6 = 180 degree pulse measured L1 = set as calculated above D5 = 5 us D3 = 5 us D4 = will be set later D6 = will be set later NE = 512 W (typically for 12 h experiment) IN = will be set later S4 = 60HNext, we need to set up the 2D parameters, which can be displayed on the screen using the ST2D command. Set TD, SI, TD2, and SI2 equal to 2K. Check to make sure TD1 is 512W if you want to take 512 experiments. Set ND0 = 2.
Now we can finish setting the parameters which were skipped over before.
Type DW to see what value it is set to (this value should be something like
90 and it is expressed in microseconds). Set IN=DW, but be careful because
DW is expressed in microseconds and IN is set in seconds, so if DW = 90,
you need to type in IN = 90u (the letter "u" stands for microseconds).
Always change IN, NOT DW. Next check the value of DE by typing DE and
hitting return. DE is also expressed in usec, and set D4 = DE - 5usec.
Next check the value of AQ, which is expressed in seconds and set
D6 = AQ + 2msec. Typical values for D4 and D6 are around 125 usec and
186 msec respectively. Finally, set DS = 2,
DR = 16 initial and 12 final, and set NS to a multiple of 16 (NS = 32
will take approximately 11-12 h).
The last thing that needs to be done before running the experiment is to set
the receiver phase (PH9). The pulse programs DANTOC.TST and DANCTOC.TST have
been created to allow you to calculate the appropriate phase. Set NS=1 and AU
the appropriate experiment. Change the receiver gain if necessary and retype AU
until your receiver gain is correctly set (PLEASE NOTE: RGA will not work! The
receiver gain must be set manually). Now restore NS to your desired value and
collect one full experiment (it should stop after one experiment). Phase the
spectrum using the "A" command (additive phase correction) in the EP mode and
type "M" to memorize the phasing. Type the command "TY PHC0" and note the value
of the zero order phase correction. Use the following formula to calculate PH9:
PH9 = 360-(PHC0+90), but do not use negative values of PH9. If PH9 turns
out negative from the above formula, add 360 to get a positive PH9. Now, enter
this value into the pulse program DANTOC.AU or DANCTOC.AU (whichever you are using).
Type "EDIT pulse program name" and move the cursor down to the line that starts out
"PH9= (360)". After the (360), enter your calculated value of PH9 so the line now reads
something like: PH9= (360) 261. Type escX to save your changes and exit the editor.
AU either DANTOC.AU or DANCTOC.AU and enter your desired
Acknowledgment: This pulse program was written by Rachel Maplestone.
Last Update: 9/27/95.
P.S. The AM500 has a tendency to crash if you leave the instrument in the job
where the 2D experiment is running. As a precaution, move to a different
experiment before leaving.
In addtion to these remarks, comments can be found at
the bottom of the .AU programs.
Questions, comments to: Andrea Sefler
andreafire.cchem.berkeley.edu /tt>